1 Scope
This Technical Specification defines terms and specifies methods for quantifying the imaging performance of real-time, ultrasound B-mode scanners. The types of transducers used (see sections 7.6 and 10.7 of [1]) with these scanners include
-
a) phased array,
-
b) linear arrays,
-
c) convex arrays,
-
d) mechanical sector scanners,
-
e) 3-D probes operating in 2-D imaging mode (see Annex K),
-
f) 3-D probes operating in 3-D imaging mode for a limited number of sets of reconstructed 2-D images (see Annex K).
The test methodology is applicable for transducers operating in the 2 MHz to 15 MHz frequency range.