1 Scope and object
This clause of Part 1 is applicable, except as follows.
1.1 Scope
1.1.1 Equipment included in scope
Replacement:
Replace the text by the following paragraphs:
This part of IEC 61010 specifies the safety requirements for electric <sc> ACTUATOR</sc>s and <sc> SOLENOID</sc> s, as applied to valves, intended to be installed in an industrial process or discrete control environment.
This part of IEC 61010 specifies:
-
— particular safety requirements for general purpose electrically operated valve <sc> ACTUATOR</sc> s and <sc> SOLENOID</sc> s,
-
— related verification tests.
The general purpose electrically operated valve <sc> ACTUATOR</sc> s and <sc> SOLENOID</sc> s, covered by this part of IEC 61010 are limited to:
-
— those rated 600 V alternative current/ 840 V direct current or less,
Service personnel interface to equipment included in the scope of this document.
1.1.2 Equipment excluded from scope
Addition at the end of the list:
This standard excludes
-
— electric <sc> ACTUATOR</sc> s and <sc> SOLENOID</sc> s for use in explosive atmospheres, as covered by the IEC 60079 series of standards;
-
— mechanical parts/aspects of valves;
-
— <sc> ACTUATOR</sc> s and <sc> SOLENOID</sc> s performing a safety function as covered by the IEC 61508 series of standards;
-
— POSITIONERs.
NOTE A positioner is defined as a "physical unit delivering an additional, often mechanical, feedback to a mechanical final controlling element that improves its velocity and precision" in IEC 60050‑351:2013, 351-56-17.