1 Scope
This document describes rules for the procedures, applications, and systems of thermoluminescence dosimetry (TLD) for dose measurements
according to the probe method. It is particularly applicable to solid “TL detectors”, i.e. rods, chips, and microcubes, made from LiF:Mg,Ti or
LiF:Mg,Cu,P in crystalline or polycrystalline form. It is not applicable to LiF powders because their use requires special procedures. The probe
method encompasses the arrangement, particularly in a water phantom or in a tissue-equivalent phantom, of single TL detectors or of “TL probes”,
i.e. sets of TL detectors arranged in thin-walled polymethyl methacrylate (PMMA) casings.
The purpose of these rules is to guarantee the reliability and the accuracy indispensable in clinical dosimetry when applied on or in the
patient or phantom. This document applies to dosimetry in teletherapy with both photon radiation from 20 keV to 50 MeV and electron radiation
from 4 MeV to 25 MeV, as well as in brachytherapy with photon-emitting radionuclides. These applications are complementary to the use of
ionization chambers.