1 Scope
This document specifies tests for the emissions of volatile organic
compounds (vocs) from the gas pathways of a medical
device, its parts or accessories, which are intended to provide
respiratory care or supply substances via the respiratory tract to a
patient in all environments. The tests of this document are intended to
quantify emissions of vocs that are added to the respirable gas stream by
the materials of the gas pathway. This document establishes acceptance
criteria for these tests.
This document addresses potential contamination of the gas stream arising from the
gas pathways, which is then conducted to the patient.
This document applies over the expected service life of the medical
device in normal use and takes into account the effects of any
intended processing or reprocessing.
This document does not address biological evaluation of the surfaces of gas
pathways that are in direct contact with the patient. The
requirements for direct contact surfaces are found in the ISO 10993 series[1].
Medical devices, parts or accessories containing gas
pathways that are addressed by this document include, but are not limited
to, ventilators, anaesthesia workstations (including gas mixers), breathing systems,
oxygen conserving devices, oxygen concentrators, nebulizers, low-pressure hose
assemblies, humidifiers, heat and moisture exchangers, respiratory gas monitors,
respiration monitors, masks, mouth pieces, resuscitators, breathing tubes, breathing
systems filters, Y-pieces and any breathing accessories intended to be used
with such devices. The enclosed chamber of an incubator, including the mattress, and
the inner surface of an oxygen hood are considered to be gas pathways and
are also addressed by this document.
This document does not address contamination already present in the gas supplied from
the gas sources while medical devices are in normal use.
EXAMPLE
Contamination arriving at the medical device from gas sources such as
medical gas pipeline systems (including the non-return valves in
the pipeline outlets), outlets of pressure regulators connected or integral to a
medical gas cylinder or room air taken into the medical device is not
addressed by ISO 18562 series.
This document is intended to be read in conjunction with ISO 18562‑1.
NOTE This document has been prepared to address the relevant essential principles of safety and performance as indicated in Annex B.